Panel Meter : Watt meter

M4Y/M4W/M5W/M4M Series

M4Y/M4W/M5W/M4M Series

 

*Max. display : 19999(M5W), 1999(Others)
*Auto Zero function and Hold function(Except for M5W)
*Display the output(0-10VDC) from Power converter.
 (It is available to correspond when output is DC4-20mA, DC1-5V.)
*7 Segment LED Display
*Case size by DIN specification.
*Indication type only, Single preset output type, Dual preset output type
Specification Table

Classification Indicator Single preset output type Dual preset output type
Measurement DC, AC voltage M4Y-DV-☐
M4Y-AV-☐
M5W-DV-☐
M5W-AV-☐
M4W-DV-☐
M4W-AV-☐-☐
M4M-DV-☐
M4M-AV-☐-☐
M4W1P-DV-☐
M4W1P-AV-☐
M4M1P-DV-☐
M4M1P-AV-☐
M4W2P-DV-☐
M4W2P-AV-☐
M4M2P-DV-☐
M4M2P-AV-☐
DC, AC ampere M4Y-DA-☐
M4Y-AA-☐
M5W-DA-☐
M5W-AA-☐
M4W-DA-☐
M4W-AA-☐-☐
M4M-DA-☐
M4M-AA-☐-☐
M4W1P-DA-☐
M4W1P-AA-☐-☐
M4M1P-DA-☐
M4M1P-AA-☐-☐
M4W2P-DA-☐
M4W2P-AA-☐-☐
M4M2P-DA-☐
M4M2P-AA-☐-☐
AC power (0-10VDC) M4Y-W-☐
M5W-W-☐
M4W-W-☐
M4M-W-☐
M4W1P-W-☐
M4M1P-W-☐
M4W2P-W
M4M2P-W
rpm, speed (0-10VDC) (0-10VAC) M4Y-T☐-☐
M4Y-S☐-☐
M5W-T-☐
M5W-S-☐
M4W-T☐-☐
M4W-S☐- ☐
M4M-T- ☐
M4M-S- ☐
M4W1P-T☐-☐
M4W1P-S☐-☐
M4M1P-T-☐
M4M1P-S-☐
M4W2P-T☐-☐
M4W2P-S☐-☐
M4M2P-T-☐
M4M2P-S☐
Power factor (DC4-20mA) - M4W-P
(Refer to L-59)
- -
Max. allowable input 150% for each input specification(At 400VAC:120%)
Power supply 100-240VAC 50/60Hz
★ 5VDC
(Except for M5W)
★ 24-70VDC
110/220VAC 50/60Hz
★ 24-70VDC
★ 100-240VAC 50/60Hz
Allowable voltage range 90 to 110% of rated voltage
Power consumption DC: 2W, AC: 4VA DC: 3W, AC: 5VA
Display method 7 Segment LED display(Charachter height: 14mm)
Character height M4Y, M4W, M5W : 14mm / M4W1P, M4W2P, M4M1P, M4M2P : 10mm
Display accuracy DC : F.S. ±0.2% rdg ±1 digit AC : F.S. ±0.5% rdg ±1 digit
Sampling period 300ms
A/D conversion method Dual slope Integral method
Response time 2sec.(0 to Max.)
Display frequency 2.5 tlmes/sec.
Contact capacity - Relay conatact output :
250VAC 3A 1c
Relay conatact output :
250VAC 3A 1c×2
Insulation resistance 100MΩ(at 500VDC megger)
Dielectric strength 2000VAC 50/60Hz for 1 minute
Noise strength ±1kV the square wave noise(pulse width : 1㎲) by the noise simulator
Vibration Mechanical 0.75mm amplitude at frequency of 10 to 55Hz(for 1 min.) in each of X, Y, Z directions for 1 hour
Malfunction 0.5mm amplitude at frequency of 10 to 55Hz(for 1 min.) In each of X, Y, Z directions for 10 minutes
Shock Mechanical 300m/s²(30G) in X, Y, Z directions for 3 times
Malfunction 100m/s²(10G) in X, Y, Z directions for 3 times
Relay life cycle Mechanical - Min. 10,000,000 operations
Malfunction - Min. 100,000 operations (250VAC 3A resistive load)
Environ-ment Ambient emperature -10 to 50℃, storage: -20 to 60℃
Ambient humidity 35 to 85%RH, storage: 35 to 85%RH
Unit weight M4Y:Approx. 144g
M5W:Approx. 172g
M4W:Approx. 168g
M4M:Approx. 262g
(M4M-P:Approx. 268g)
M4W1P:Approx. 253g
M4M1P:Approx. 290g
M4W2P:Approx. 278g
M4M2P:Approx. 316g

※ "★" symbol in power supply is optional. (Customizable)

 