Percent Measurement

Model 201RS Panel Mount Percent Oxygen analysers with Complete Sample System

 

  • High resolution 3 ½ LCD
  • Comprehensive state-of-the-art electronics platform, includes: simple and intuitive user interface, 2-fully adjustable alarm relay contacts, programmable alarm configurations, alarm bypass feature, programmable alarm delays, on board data logger, 3-individual security settings, internal diagnostics, remaining sensor life indication, etc.
  • Standard analog output ranges: 0-1%, 0-5%, 0-10% and 0-25%
  • Optional analog output ranges: 0-1000ppm, 0-50% or 0-100%
  • Isolated 4-20mA. analog output
  • Patented cellblock technology. Provides a flow control valve, flow meter, internal span gas and isolation solenoid valves all integral to the compact cell block
    General Purpose Area Classification

 

 

Model 210BR Wall Mount or 2.0" pipe Mount Hazardous Location Percent Oxygen analysers With Complete Sample System

 

  • High resolution 3 ½ digit LCD
  • Standard analog output ranges: 0-1%, 0-5%, 0-10% and 0-25%
  • Optional analog output ranges: 0-1000ppm, 0-50% or 0-100%
  • 1-5VDC and 4-20mA. analog outputs
  • Patented cellblock technology. Provides a flow control valve, 3-way selector valve: Sample/Span/Off, and flow meter, all integral to the compact cell block
  • Approved by CSA International to UL and CSA standards as meeting requirements for Class I, Div. 1, Groups B,C,D with flammable gas sample

 

Model 70 Panel Mount Percent Oxygen Analyser

 

  • High resolution 3 ½ digit LCD
  • Standard analog output ranges: 0-1%, 0-5%, 0-10% and 0-25%
  • Optional analog output ranges: 0-50%, or 0-95%
  • 0-1VDC and isolated 4-20mA. analog outputs
  • Remarkably stable 10-year life zirconium oxide sensor
  • 2- fully adjustable alarm relay contacts
  • General Purpose Area Classification

 

Model 111 Series Portable Percent Oxygen analysers

  • High resolution 3 ½ digit LCD
  • Standard analog output ranges: 0-1%, 0-5%, 0-10% and 0-25%
  • Optional analog output ranges: 0-50%, or 0-95%
  • On board data logger
  • Remarkably stable 10-year life zirconium oxide sensor
  • Flow control valve and flow meter integrated into advanced technology cellblock
  • The Model 111 has been designed for positive pressure applications
  • The Model 111P and 111B are designed for negative pressure to 5psig. applications and are outfitted with long life sample pumps
  • All have Rechargeable NiCad batteries
  • The Model 111 and Model 111P are for use in General Purpose Area Classifications
  • The Model 111B has been designed to meet Class1, Div. 2, Groups C,D Area Classifications